The IS6-D-IR-170 calibrated integrating spheres is 5.3 in. inside diameter with a Germanium detector for use with highly divergent beams up to 170°. The IR detector is calibrated from 700 to 1800 nm and can measure optical power up to 30 W.
- 5.3 in. ID 4-port Integrating sphere for up to 170° divergent beams
- Germanium photodetector with filter for 700-1800 nm spectral range
- 20 µW to 30 W power measurement range
- Port plugs, covers, adapters and reducers available See All Features
Specifications
- Product NameIS6-D-IR-170
- TypeDivergent Beam
- Sphere Size5.3 inch Inner Diameter
- Aperture SizeØ8 mm
- Detector TypeGermanium
- Spectral Range700-1800 nm
- Minimum Power20 µW
- Maximum Average Power30 W
- Maximum Pulse Energy0.7 mJ
- Maximum Average Power Density1 kW/cm²
- Maximum Beam Divergence±85 deg
- Beam Divergence Sensitivity±1.5%
- CE ComplianceYes
- UKCA ComplianceYes
- China RoHS ComplianceYes
Features
Integrating Spheres for Divergent Light Sources
Integrating spheres are used when we have divergent light sources. An integrating sphere has its inner surface coated with a surface that highly reflects (typically 99%) in a scattering, nonspecular way. Thus when a divergent beam hits the walls of the integrating sphere, the light is reflected and scattered many times until the light hitting any place on the walls of the sphere has the same intensity.
A detector placed in the sphere thus gets the same intensity as anywhere else and the power the detector detects is thus proportional to the total incident power independent of the beam divergence. (The detector is so arranged that it only sees scattered light and not the incident beam). An ideal integrating sphere has a surface with reflective properties that are Lambertian. This means that light incident on the surface is scattered uniformly in all directions in the 2pi steradians solid angle above the surface. The surface used by Ophir closely approximates a Lambertian surface.
IS6 Integrating Sphere Overview
For applications that need a large Integrating Sphere, Ophir offers the IS6 series. These are 6” Integrating Spheres, available with and without built-in calibrated sensors, in a range of configurations. Get to know the IS6 family in this video.
Measuring Beams Coming Out of A Fiber
When you need to measure a beam coming out of a fiber, there are some parameters that might have a somewhat different meaning than they do when referring to "regular" beam measurements. Missing some of these points could lead to incorrect measurement, and possible equipment damage. This video clarifies some issues you'll need to keep in mind so you can set up -- and perform -- this measurement correctly.
Measuring Power of LEDs: UV, Visible and NIR
Measuring the emitted power of an LED can be tricky. It is different in some important ways from measuring the power of a laser beam. This video shows you how to use the Ophir 3A-IS Integrating Sphere Sensor, along with the Auxiliary LED accessory, to easily make accurate measurements in LED applications.
Calibration Factors - Laser Power/Energy Meter
When a power/energy meter is in "Calibrate" mode, various "Factors" are displayed to the user. This video explains the meaning of each of these factors.
IS6 Integrating Sphere Selection Guide
| 200-1100 nm 300nw-1W |
400-1100 nm 20µW-30W |
700-1800 nm 20µW-30W |
|
|---|---|---|---|
| Collimated beam < ±15° Beam diameter < 25mm |
IS6-C-UV | IS6-C-VIS | IS6-C-IR |
| Collimated beam < ±30° Beam diameter > 25mm |
IS6-C-UV-2.5’’ | ||
| Divergent beam > ±15° Max divergence < ±40° Beam diameter < 25mm |
IS6-D-UV | IS6-D-VIS | IS6-D-IR |
| Divergent beam > ±15° Max divergence < ±56° Beam diameter < 10mm |
|||
| Divergent beam > ±15° Max divergence < ±60° Beam diameter < 5mm |
|||
| Highly divergent beam <±85° Beam diameter < 8mm |
IS6-D-IR-170 |
Frequently Asked Questions
- There are several models of IS6 integrating sphere detectors. how can I select the right one?Answer
There is a simple to use selection guide.
- Integrating spheres are used when you have divergent light sources. How do they work?Answer
Integrating Sphere Theory
Integrating spheres are used when we have divergent light sources. As shown in the illustration, an integrating sphere has its inner surface coated with a surface that highly reflects (typically 99%) in a scattering, nonspecular way. Thus when a divergent beam hits the walls of the integrating sphere, the light is reflected and scattered many times until the light hitting any place on the walls of the sphere has the same intensity.A detector placed in the sphere thus gets the same intensity as anywhere else and the power the detector detects is thus proportional to the total incident power independent of the beam divergence. (The detector is so arranged that it only sees scattered light and not the incident beam). An ideal integrating sphere has a surface with reflective properties are Lambertian. This means that light incident on the surface is scattered uniformly in all directions in the 2pi steradians solid angle above the surface. The surface used by Ophir closely approximates a Lambertian surface.

Step 1 – Starting position3A-IS Series
The 3A-IS series has two 50mm integrating spheres in series with a photodiode detector. The two series spheres scramble up the light very well thus giving output very independent of incident beam divergence angle. The two spheres in series also insure that the light hitting the detector is greatly reduced in intensity thus allowing use up to 3 Watts even though photodiodes saturate at about 1mW. There are two models, the 3A-IS with a silicon photodiode for 400 – 1100nm and the 3A-ISIRG with an InGaAs detector for 800 – 1700nm
- Are there any special problems with the calibration stability of integrating sphere sensors?Answer
The Ophir integrating sphere sensors, models 3A-IS and 3A-IS-IRG have a white diffuse reflecting coating on the inside of the integrating sphere. The sensitivity of the sensor is quite sensitive to the reflectivity of the coating. If the coating absorption goes up 1%, it can cause a 5% change in reading. Therefore, care must be taken not to soil or damage the white coating of the sensors. Also it may be a good idea to send the sensors for recalibration yearly.
- When using the fiber optic adaptor, how do we handle power loss due to the fiber relative to calibration?Answer
All Ophir power meters, including photodiode power meters, have an air gap between the fiber tip and the sensor. Therefore they measure the power emitted by the fiber into the air and do not take into account any reflection losses there are in the fiber. Therefore, if in actual use, the fiber will be coupled with no loss to another element, then the losses should be added to the reading. These losses are usually about 4%. Thus if the reading on the Ophir meter is say 100 mW, then in lossless use, the real power will be 104 mW.
- Among the Integrating Sphere accessories offered, there are “Port Plugs” (white), and “Port Covers” (black). What’s the difference?Answer
An unused port should be closed, to prevent unwanted light from entering the sphere. Closing it with a diffuse white port plug, however, adds the surface area of that plug to the (diffuse white) effective area of the sphere that is doing the “integrating”. For a calibrated integrating sphere sensor, this change in the behavior of the sphere changes its calibration, and results in incorrect readings. In such applications, a black “Port Cover” should be used.
- The IS6 integrating spheres have a specified “Sensitivity to beam size” and “Sensitivity to beam divergence”. What is that?Answer
In general, as the divergence angle of the beam entering the integrating sphere increases - and as its diameter increases – the assumptions on which we base the sphere’s performance (infinite reflections inside the sphere walls, perfectly uniform distribution of light inside the sphere, etc.) become less correct. We therefore specify the maximum beam divergence (such as ± 60⁰), and we also state the maximum possible change in reading caused by change in beam size. In fact, we also state in the data sheet that the maximum additional uncertainty due to beam size is only ±1% for beam divergence < 30⁰, and ±3% for beam divergence > 30⁰. To give this more meaning: Basically, if you measure the power using a beam that is a few mm in diameter, that has a relatively small divergence angle, and is centered on the sphere’s input port aperture, you can safely ignore this additional uncertainty.
- Do I need to recalibrate my instrument? How often must it be recalibrated?Answer
Unless otherwise indicated, Ophir sensors and meters should be recalibrated within 18 months after initial purchase, and then once a year after that.
- Can I use my IS6-D Integrating Sphere (which is normally used for measuring Divergent beams) to measure a Collimated beam? I know that normally one would use an IS6-C for Collimated beams, but can I manage with my -D sphere on a 1-time basis?Answer
Here is a trick that would make this possible:
The beam should be aimed so that it is incident close to the detector port (but not hitting the baffle) – as shown in this drawing:
This way the "first bounce" will be directed to the opposite side of the sphere, ensuring that the detector will in fact see only light from the "second bounce" and onward, i.e. light that has been uniformly distributed around the inner sphere surface (normally, light from the "first bounce" of a collimated beam is not yet uniformly distributed and we don’t want the detector to see it – that is the main idea behind the different C and D configurations. This trick gets around that).
Accessories
Integrating Sphere Accessories
| Compare | Description | Drawings, CAD & Specs | Avail. | Price | ||
|---|---|---|---|---|---|---|
![]() | 7Z08283APort Plug, IS6 Integrating Sphere, Ø63.5 mm, PTFE | |||||
![]() | 7Z08280APort Plug, IS6 Integrating Sphere, Ø25.4 mm, PTFE | |||||
![]() | 7Z08281APort Cover, IS6 Integrating Sphere, Ø63.5 mm, Matte Black Coated | |||||
![]() | 7Z08282APort Cover, IS6 Integrating Sphere, Ø25.4 mm, Matte Black Coated | |||||
![]() | 7Z08305APort Reducer, 2.5 to 1 inch, IS6 Integrating Sphere, Ø63.5 mm, White Coated | |||||
![]() | 7Z08306Dovetail Flange Attachment, IS6 Integrating Sphere, for 7Z08305A Port Reducer | |||||
![]() | 7Z08307Aperture Set, Ø5, Ø7, Ø10 mm, IS6 Integrating Sphere, for 7Z08305A Port Reducer | |||||
![]() | 7Z08286FC Fiber Adapter, IS6 Integrating Sphere, Ø25.4 mm, Matte Black Coated | |||||
![]() | 7Z08350FPD Sensor Adapter, IS6 Integrating Sphere, North Pole Port, Matte Black Coated | |||||
![]() | 7Z08289SM1 Thread Adapter, IS6 Integrating Sphere, Ø25.4 mm, Matte Black Coated | |||||
![]() | 7Z08285SMA Fiber Adapter, IS6 Integrating Sphere, Ø25.4 mm, Matte Black Coated |
Extended Warranty for Sensor
Customers that purchase the above items also consider the following items. Ophir-Spiricon meters and sensors include a standard manufacturers warranty for one year. Add a one year Extended Warranty to your meter or sensor, which includes one recalibration.
| Compare | Description | Drawings, CAD & Specs | Avail. | Price | ||
|---|---|---|---|---|---|---|
![]() | XWAR-SENSORExtended Warranty for Sensor |
Resources
Data Sheets
Integrating Spheres Large Dimensions 5.3” Datasheet(700.9 kB, PDF)
Catalogs
Power Sensors Catalog(11.2 MB, PDF) Laser Power & Energy Measurement and Laser Beam Analysis Catalog(27.5 MB, PDF)
Application Notes
Drawings & CAD
IS6-C / IS6-D Drawing(33.4 kB, PDF)
Technical Notes
IS6 Sensor User Notes(1 MB, PDF) 5 Situations Where Laser Performance Measurement is Necessary How do I know what range, or scale, to set my power/energy meter to? Measuring Average Power of Pulsed Lasers with Photodiodes(114.5 kB, PDF) Measuring Power of Divergent Beams with Integrating Sphere Sensors
Technical Articles
VCSEL Measurement Solutions Ophir Power/Energy Meter Calibration Procedure and Traceability/Error Analysis Laser Measurements in Materials Processing Low Frequency Power Mode Measuring Average Power of Pulsed Lasers with Photodiodes