The Comet 10K-HD high damage threshold portable power puck laser sensor features a 55 mm aperture. It can measure power from 0.2 to 10 kW by sampling a 30 s exposure of laser power. It has a deflecting cone and the spectrally flat broadband coating and covers the spectral range 980 to 1070 nm and 10.6 µm.
- Measure heat rise from 30 s exposure to laser
- Accurate, built in temperature compensation algorithm
- Up to 10 kW power measurement
- 55 mm diameter aperture
- The Comet 10K-HD sensor is not under ISO/IEC 17025:2017 accreditation See All Features
Specifications
- Product NameComet 10K-HD
- Spectral Range0.98-1.07 µm and 10.6 µm
- Aperture SizeØ55 mm
- Minimum Power200 W
- Maximum Average Power10,000 W
- Maximum Average Power Density2 kW/cm²
- Maximum Energy Density <100 ns1 J/cm²
- Maximum Energy Density 2 ms45 J/cm²
- Dimensions355 x 87 x 110 mm
- CE ComplianceYes
- UKCA ComplianceYes
- China RoHS ComplianceYes
Features
Meet the Comet family of Laser Power Probes
Meet the Comet family of Laser Power Probes -- accurate, low cost meters that are self-contained, and fast and easy to use in industrial settings.
Choosing a Thermal Sensor
In this short “Basics” video we review the use – and selection - of thermal sensors for measuring low, medium and high laser powers.
Thermal Power Sensor Accuracy
Ophir former CTO Dr. Ephraim Greenfield discusses the various factors that contribute to uncertainties in measurement when using Ophir laser power and energy meters.
Effects of Incidence Angle on Measurements
In this video, you will learn to what degree a beam’s incidence angle matters, and for which sensor types this should be taken into account.
Frequently Asked Questions
- How should I clean my sensor?Answer
First, clean the absorber surface with a tissue, using Umicore #2 Substrate Cleaner, acetone or methanol. Then dry the surface with another tissue. Please note that a few absorbers (Pyro-BB, 10K-W, 15K-W, 16K-W and 30K-W) cannot be cleaned with this method. Instead, simply blow off the dust with clean air or nitrogen. Don't touch these absorbers. Also, HE sensors (such as the 30(150)A-HE-17) should not be cleaned with acetone.
Note: These suggestions are made without guarantee. The cleaning process may result in scratching or staining of the surface in some cases and may also change the calibration. - What is the damage threshold of the thermal broadband coating and how does it compare with claims of other manufacturers?Answer
We publish a nominal damage threshold for most of our thermal BB sensors as 20 KW/cm2. Other manufacturers may quote higher numbers than this. In actuality, in one to one tests against competitors, our sensors show a higher damage threshold but the actual damage threshold depends on the total power as well as the power density. For very low powers such as 30 W, the damage threshold can be as high as 50 KW/cm2 and at high powers such as 5 KW, it drops to 3 KW/cm2. The Ophir sensor finder program takes account of these variations in its calculations.
- Can a laser measurement depend on the distance from the laser to the sensor?Answer
In theory, if a beam is completely parallel and fits within the aperture of a sensor, then it should make no difference at all what the distance is. It will be the same number of photons (ignoring absorption by the air, which is negligible except in the UV below 250nm). If, nevertheless, you do see such a distance dependence, there could be one of the following effects happening:
- If you are using a thermal type power sensor, you might actually be measuring heat from the laser itself. When very close to the laser, the thermal sensor might be “feeling” the laser’s own heat. That would not, however, continue to have an effect at more than a few cm distance unless the light source is weak and the heat source is strong.
- Beam geometry – The beam may not be parallel and may be diverging. Often, the lower intensity wings of the beam have greater divergence rate than the main portion of the beam. These may be missing the sensor's aperture as the distance increases. To check that you'd need to use a profiler, or perhaps a BeamTrack PPS (Power/Position/Size) sensor.
- If you are measuring pulse energies with a diffuser-based pyroelectric sensor: Some users find that when they start with the sensor right up close to the laser and move it away, the readings drop sharply (typically by some 6%) over the first few cm. This is likely caused by multiple reflections between the diffuser and the laser device, which at the closest distance might be causing an incorrectly high reading. You should back off from the source by at least some 5cm, more if the beam is not too divergent.
Needless to say, it’s also important to be sure to have a steady setup. A sensor held by hand could easily be moved around involuntarily, which could cause partial or complete missing of the sensor’s aperture at increasing distance, particularly for an invisible beam.
- Do I need to recalibrate my instrument? How often must it be recalibrated?Answer
Unless otherwise indicated, Ophir sensors and meters should be recalibrated within 18 months after initial purchase, and then once a year after that.
Resources
Data Sheets
Drawings & CAD
Comet 10K-HD Drawing(140.6 kB, PDF)
Manuals
Comet User Manual(94.1 kB, PDF)
Catalogs
Power Sensors Catalog(11.2 MB, PDF) Laser Power & Energy Measurement and Laser Beam Analysis Catalog(27.5 MB, PDF)
Technical Notes
5 Situations Where Laser Performance Measurement is Necessary Effect of Ambient Conditions on Laser Measurements How do I know what range, or scale, to set my power/energy meter to? Setting Your Thermal Sensor up in Energy Mode for the Best Accuracy and Repeatability How Much of the Power Sensor Aperture Can My Laser Fill Up? How to Properly Select a Laser Power or Energy Sensor Common Reasons for Thermal Sensor Damage or Out of Tolerance Conditions