The 3A-IS-IRG integrating sphere laser measurement sensor with InGaAs photodiode is for divergent beams up to +/-40 degrees. It has a 12 mm aperture and can measure optical power from 1 µW to 3 W. It covers the spectral range from 800 to 1700 nm.
- 1.6 inch ID integrating sphere for divergent beam
- InGaAs photodiode for 800 to 1700 nm spectral range
- 1 µW to 3 W power measurement range
- 12 mm diameter aperture See All Features
Specifications
- Product Name3A-IS-IRG
- TypeDivergent Beam
- Sphere Size1.6 inch Inner Diameter
- Aperture SizeØ12 mm
- Detector TypeInGaAs Photodiode
- Spectral Range800-1700 nm
- Minimum Power1 µW
- Maximum Average Power3 W
- Maximum Pulse Energy500 µJ
- Maximum Average Power Density0.2 kW/cm²
- Response Time0.2 s
- Dimensions68 x 95 x 46 mm (LxWxD)
- Cable Length1.5 m
- CE ComplianceYes
- UKCA ComplianceYes
- China RoHS ComplianceYes
Features
Integrating Spheres for Divergent Light Sources
Integrating spheres are used when we have divergent light sources. An integrating sphere has its inner surface coated with a surface that highly reflects (typically 99%) in a scattering, nonspecular way. Thus when a divergent beam hits the walls of the integrating sphere, the light is reflected and scattered many times until the light hitting any place on the walls of the sphere has the same intensity.
A detector placed in the sphere thus gets the same intensity as anywhere else and the power the detector detects is thus proportional to the total incident power independent of the beam divergence. (The detector is so arranged that it only sees scattered light and not the incident beam). An ideal integrating sphere has a surface with reflective properties that are Lambertian. This means that light incident on the surface is scattered uniformly in all directions in the 2pi steradians solid angle above the surface. The surface used by Ophir closely approximates a Lambertian surface.
Two Integrating Spheres in Series
The 3A-IS series has two 50 mm integrating spheres in series with a photodiode detector. The two series spheres scramble up the light very well thus giving output very independent of incident beam divergence angle. The two spheres in series also insure that the light hitting the detector is greatly reduced in intensity thus allowing use up to 3 Watts even though photodiodes saturate at about 1 mW. There are two models, the 3A-IS with a silicon photodiode for 400-1100 nm and the 3A-ISIRG with an InGaAs detector for 800-1700 nm.
Calibration Stability of 3A-IS Integrating Sphere Sensors
The Ophir integrating sphere sensors, models 3A-IS and 3A-IS-IRG have a white diffuse reflecting coating on the inside of the integrating sphere. The sensitivity of the sensor is quite sensitive to the reflectivity of the coating. If the coating absorption goes up 1%, it can cause a 5% change in reading. Therefore, care must be taken not to soil or damage the white coating of the sensors. Also it may be a good idea to send the sensors for recalibration yearly.
How to Measure Laser Power in Telecom Applications
Optical laser measurements in telecom applications have unique challenges. This video will introduce you to the solutions Ophir offers to help you.
Measuring Beams Coming Out of A Fiber
When you need to measure a beam coming out of a fiber, there are some parameters that might have a somewhat different meaning than they do when referring to "regular" beam measurements. Missing some of these points could lead to incorrect measurement, and possible equipment damage. This video clarifies some issues you'll need to keep in mind so you can set up -- and perform -- this measurement correctly.
Frequently Asked Questions
- Are there any special problems with the calibration stability of integrating sphere sensors?Answer
The Ophir integrating sphere sensors, models 3A-IS and 3A-IS-IRG have a white diffuse reflecting coating on the inside of the integrating sphere. The sensitivity of the sensor is quite sensitive to the reflectivity of the coating. If the coating absorption goes up 1%, it can cause a 5% change in reading. Therefore, care must be taken not to soil or damage the white coating of the sensors. Also it may be a good idea to send the sensors for recalibration yearly.
- Integrating spheres are used when you have divergent light sources. How do they work?Answer
Integrating Sphere Theory
Integrating spheres are used when we have divergent light sources. As shown in the illustration, an integrating sphere has its inner surface coated with a surface that highly reflects (typically 99%) in a scattering, nonspecular way. Thus when a divergent beam hits the walls of the integrating sphere, the light is reflected and scattered many times until the light hitting any place on the walls of the sphere has the same intensity.A detector placed in the sphere thus gets the same intensity as anywhere else and the power the detector detects is thus proportional to the total incident power independent of the beam divergence. (The detector is so arranged that it only sees scattered light and not the incident beam). An ideal integrating sphere has a surface with reflective properties are Lambertian. This means that light incident on the surface is scattered uniformly in all directions in the 2pi steradians solid angle above the surface. The surface used by Ophir closely approximates a Lambertian surface.

Step 1 – Starting position3A-IS Series
The 3A-IS series has two 50mm integrating spheres in series with a photodiode detector. The two series spheres scramble up the light very well thus giving output very independent of incident beam divergence angle. The two spheres in series also insure that the light hitting the detector is greatly reduced in intensity thus allowing use up to 3 Watts even though photodiodes saturate at about 1mW. There are two models, the 3A-IS with a silicon photodiode for 400 – 1100nm and the 3A-ISIRG with an InGaAs detector for 800 – 1700nm
- When using the fiber optic adaptor, how do we handle power loss due to the fiber relative to calibration?Answer
All Ophir power meters, including photodiode power meters, have an air gap between the fiber tip and the sensor. Therefore they measure the power emitted by the fiber into the air and do not take into account any reflection losses there are in the fiber. Therefore, if in actual use, the fiber will be coupled with no loss to another element, then the losses should be added to the reading. These losses are usually about 4%. Thus if the reading on the Ophir meter is say 100 mW, then in lossless use, the real power will be 104 mW.
- When measuring a fiber output, should I put the fiber tip right up against the PD300 detector? If not, how close should I come?Answer
The PD300 sensors are not designed to measure with the fiber pushed up right against the detector surface. It may be reading lower in such a case due to saturation of the detector from the concentrated beam or higher due to back reflections off the detector and back again from the fiber tip. The optimal reading will be where the beam is expanded to a size of 2-5mm diameter. Therefore, you should back off the fiber to a distance where the beam has expanded somewhat. Do not back off too far, otherwise if the nominal beam size is larger than given above, you may lose some of the beam off the edges of the detector.
- Can the PD300 be used with the filter in and with fiber adapter at the same time?Answer
Yes, the adapter is arranged so the user can use it either with filter out or in.
- Among the Integrating Sphere accessories offered, there are “Port Plugs” (white), and “Port Covers” (black). What’s the difference?Answer
An unused port should be closed, to prevent unwanted light from entering the sphere. Closing it with a diffuse white port plug, however, adds the surface area of that plug to the (diffuse white) effective area of the sphere that is doing the “integrating”. For a calibrated integrating sphere sensor, this change in the behavior of the sphere changes its calibration, and results in incorrect readings. In such applications, a black “Port Cover” should be used.
- Do I need to recalibrate my instrument? How often must it be recalibrated?Answer
Unless otherwise indicated, Ophir sensors and meters should be recalibrated within 18 months after initial purchase, and then once a year after that.
- How much variation in water temperature or flow rate is allowed in Ophir water cooled sensors?Answer
Ophir water cooled sensors measure the heat flow across the thermopile disc and therefore are quite insensitive to the water temperature or flow rate within the given specified limits. However, sudden changes in the water temperature or water flow rate can cause a disturbance to the reading until the flow rate/temperature stabilizes again. Therefore we specify in our water cooled sensors that the water temperature should not change faster than 1 °C/min. Likewise, sudden changes in flow rate (e.g. switching another device connected to the same water line on and off) can results in temporary disturbances in the power reading.
- There are several models of IS6 integrating sphere detectors. how can I select the right one?Answer
There is a simple to use selection guide.
Accessories
Fiber Connector Adapters
These adapters allow for power measurement of connectorized fiber-optic cables. The sensor may need an additional mounting bracket to connect to these fiber adapters.
| Compare | Description | Drawings, CAD & Specs | Avail. | Price | ||
|---|---|---|---|---|---|---|
![]() | 7Z08227SC Fiber Connector Adapter, Power and Energy Sensors | |||||
![]() | 7Z08226ST Fiber Connector Adapter, Power and Energy Sensors | |||||
![]() | 7Z08229FC Fiber Connector Adapter, Power and Energy Sensors | |||||
![]() | 1G01236ASMA Fiber Connector Adapter, Power and Energy Sensors |
Fiber Adapter Mounting Bracket
A mounting bracket is needed to connect most power and energy sensors to a fiber adapter (SC, ST, FC or SMA). This bracket can be used for integrating sphere models 3A-IS and 3A-IS-IRG.
| Compare | Description | Drawings, CAD & Specs | Avail. | Price | ||
|---|---|---|---|---|---|---|
![]() | 7Z08213Fiber Adapter Mounting Bracket, 3A-IS Photodiode Sensors |
Extended Warranty for Sensor
Customers that purchase the above items also consider the following items. Ophir-Spiricon meters and sensors include a standard manufacturers warranty for one year. Add a one year Extended Warranty to your meter or sensor, which includes one recalibration.
| Compare | Description | Drawings, CAD & Specs | Avail. | Price | ||
|---|---|---|---|---|---|---|
![]() | XWAR-SENSORExtended Warranty for Sensor |
Resources
Data Sheets
Integrating Spheres Small Dimensions 1”-1.6” Datasheet(171.1 kB, PDF)
Catalogs
Power Sensors Catalog(11.2 MB, PDF) Laser Power & Energy Measurement and Laser Beam Analysis Catalog(27.5 MB, PDF)
Drawings & CAD
3A-IS-IRG Drawing(156.6 kB, PDF)
Application Notes
Tutorials
Technical Notes
Measuring Power of Divergent Beams with Integrating Sphere Sensors 5 Situations Where Laser Performance Measurement is Necessary How do I know what range, or scale, to set my power/energy meter to? Measuring Average Power of Pulsed Lasers with Photodiodes(114.5 kB, PDF)













